[an error occurred while processing this directive]
血液净化中心管理

百特金宝AK96透析机浓缩液相关故障的分析与处理

展开
  • 1. 陕西省中医医院血液透析室

收稿日期: 2016-06-27

  修回日期: 2016-11-14

  网络出版日期: 2017-05-19

Analysis and treatment of concentrate related failures of Baxter Gambro AK96 hemodialysis machine

Expand

Received date: 2016-06-27

  Revised date: 2016-11-14

  Online published: 2017-05-19

摘要

通过对百特金宝AK96 血液透析机“碳酸氢盐使用不正确,检查浓缩液”这一故障进行简单描述,深入的分析了设备运行原理,提出了处理百特金宝AK96 透析机浓缩液相关故障的方案,该方案具有实际可操作性,可为医院透析行业工程技术人员及血透机的操作者提供参考。

本文引用格式

谷林,于小勇,陈文静,冯亚红 . 百特金宝AK96透析机浓缩液相关故障的分析与处理[J]. 中国血液净化, 2017 , 16(05) : 350 -352 . DOI: 10.3969/j.issn.1671-4091.2017.05.018

Abstract

This article briefly describes the fault of“Incorrect bicarbonate concentrate, check concentrate” demonstrated on Baxter Gambro AK96 Hemodialysis Machine, analyzes the operation principle of the equipment in details, and then proposes the treatment schemes about this fault, which have practical operability, and can provide reference for technologists and operators working in dialysis department.
文章导航

/

[an error occurred while processing this directive]