Comparison of flow rate in different bloodlines in dialyzers under negative pressure

  • MA Ying-Hui ,
  • WANG Yan ,
  • YIN Jia-Ning ,
  • LIU Jing
Expand
  • 1Blood Purification Center, Youan Hospital, Capital Medical University,Beijing 100069, China;  2Renal Division, Peking University First Hospital, Beijing 100034, China;  3Department of Blood Purification, Beijing Chao-Yang Hospital, Capital Medical University, Beijing 100020, China; 4Institute of Nephrology, Capital Medical University, Beijing 100069, China

Received date: 2019-03-29

  Revised date: 2019-06-28

  Online published: 2019-09-27

摘要

【摘要】目的   比较泵前存在负压的条件下不同血液透析管路的实际流速。方法   选取4种不同品牌的管路A、B、C、D,分别以200ml/min和300ml/min的设定流速运转5min,测量不同泵前负压条件下各管路的实际流速。结果①血泵设定流速相同时,泵前负压越大实际流速越低;②设定转速为200ml/min 时,- 120mmHg(1mmHg=0.133kpa)条件下不同品牌管路的实际流速无统计学差异(F =2.095,P =0.120),-100mmHg 和-200mmHg 条件下不同品牌管路的实际流速存在统计学差异(F =4.157, 11.128,P=0.014, <0.001);③设定转速为300ml/min 时, -100mmHg 和-200mmHg 条件下不同品牌管路的实际流速存在统计学差异(F=6.861, 18.748, P=0.001,<0.001)。结论泵前负压影响体外循环的实际流速,且不同品牌的管路在相同负压条件下实际流速有一定差异,可进一步研究该差异对临床治疗效果的影响。

关键词: 血液透析; 负压; 管路; 流速

本文引用格式

马颖慧 , 王艳 , 尹佳宁 , 刘静 . 负压条件下不同血液透析管路的流速比较[J]. 中国血液净化, 2019 , 18(10) : 697 -700 . DOI: 10.3969/j.issn.1671-4091.2019.10.010

Abstract

【Abstract】Objective To compare the flow rate in bloodlines of different dialyzer under negative pressure. Methods In this paper, bloodlines A, B, C and D in 4 brands of dialyzers were selected. Each bloodline ran at the speed of 200 ml/min and 300 ml/min for 5 minutes, and the actual flow rate in each bloodline was measured under different pre-pump negative pressures. Results ①The greater the pre-pump negative pressure was, the lower the actual flow rate became under the same rate setting. ②There was no statistical difference in actual flow rate in different brands of bloodlines under the set conditions of flow rate 200 ml/min and the pressure -20mmHg (F=2.095, P=0.120). In contrast, there was statistical difference in actual flow rate in
the bloodlines under the pressure of -100mmHg and -200mmHg (F=4.157 and 11.128 respectively; P=0.014 and <0.001 respectively). ③There was statistical difference in actual flow rate in different brands of bloodlines under the rate of 300ml/min and the pressure of -100mmHg and -200mmHg (F=6.861 and 18.748 respectively; P=0.001 and <0.001 respectively). Conclusion The negative pre-pump pressure affected the actual flow rate of extracorporeal circulation. The difference of actual flow rate existed among bloodlines of different brands under the same negative pressure conditions. Therefore, the clinical effect of these differences should be further studied.

参考文献

[1] 梅长林,叶朝阳,赵学智. 实用透析手册[M]. 北京: 人民卫生出版社,2004: 166.
[2] 徐英, 赵智芳, 詹云凤, 等. 增加超滤量及血流量对血液透析充分性的影响[J]. 中国临床研究, 2014; 27(1): 41-42.
[3] HD. Polaschegg. The extracorporeal circuit. Seminars in Dialysis, 1995(8), 5:299-304.
[4] 陈香美.血液净化标准操作规程(2010版)[M]. 北京: 人民军医出版社, 2010; 54-55.
[5] YY 0267-2016 血液透析及相关治疗 血液净化装置的体外循环管路.
[6] 高慧莹, 刘涛, 孙振杰. 蠕动泵原理及在化学机械抛光过程中的应用[J]. 电子工业专用设备, 2010; (9): 48-51.
[7] 王道臣, 陈志军, 韩玉明, 等. 蠕动泵流量的理论计算与试验验证[J]. 化工自动化及仪表, 2015; (2): 186-188.
[8] 马颖慧, 刘静. 不同血液透析管路的流速比较. 中国医学装备协会第二十六届学术与技术交流年会论文汇编[C]. 2017:24-25.
[9] 李文秀, 温玉, 潘玮瑛, 等. 血液流速和穿刺法对血液透析患者透析效果和内瘘血管的影响[J]. 中国医学创新, 2013; 28(10): 6-8.
[10] 刘长万. 老年患者血液透析中血流量不足的影响因素与对策[J]. 临床合理用药, 2011; 4(2): 30-31.
[11] L Yang, A Yin, W Liu. Variation of flow rate and angle of injected venous needle on influencing intimal hyperplasia at the venous anastomosis of the hemodialysis graft[J]. Australas Phys Eng Sci Med, 2017; 40: 239.
[12] 牛莉. 透析血流量对透析效果及患者生存质量的影响. 硕士论文, 吉林大学. 2015.
[13] Kyung YC, Su-Hyun K, Young OK, et al. The impact of blood flow rate during hemodialysis on all-cause mortality[J]. Korean J Intern Med, 2016; 31:1131-1139.
文章导航

/

[an error occurred while processing this directive]