A comparative study of two different types of dialysis machines

  • ZHAO Xiao-Lin ,
  • REN Qin-Qin ,
  • ZHU Han-Yu ,
  • MA Zhi-Fang
Expand
  • 1Department of Nephrology, First Medical Center of PLA General Hospital, Beijing 100853, China

Received date: 2018-07-30

  Revised date: 2019-08-22

  Online published: 2019-10-28

摘要

【摘要】目的对比中央透析液供给系统(central dialysis fluid delivery system,CDDS)与普通透析机2 种类型设备的透析液洁净度以及故障率,分析CDDS 透析设备的临床应用优势与不足。方法于2017 年8 月~2018 年3 月选择解放军总医院透析中心CDDS 透析机GC-110N 与贝朗德佳透析机各5 台,对比2 组透析机入口与出口处样本细菌、内毒素水平。回顾性分析解放军总医院2 类透析设备各18 台在2017 年3 月~2019 年3 月期间水路、电路以及其他故障的发生率。结果透析机入口采样点CDDS 组透析机细菌、内毒素水平均低于普通组透析机(t 值分别为11.215、6.468,P 值分别为<0.001,<0.001)。透析机出口采样点CDDS 组透析机细菌、内毒素水平均低于普通组透析机(t 值分别为9.154、5.490,P 值分别为<0.001,<0.001)。CDDS 组透析机水路、电路以及其他故障的总计发生率均低于普通组透析机, 差异有统计学意义(t=9.147,P=0.010)。结论CDDS 系统可提高透析液洁净度,降低透析设备故障率在临床应用中具有一定优势。

本文引用格式

赵小淋 , 任琴琴 , 朱晗玉 , 马志芳 . 2 种不同类型透析机的对比研究[J]. 中国血液净化, 2019 , 18(11) : 801 -804 . DOI: 10.3969/j.issn.1671-4091.2019.11.019

Abstract

【Abstract】Objective To compare the cleanliness and failure rate between the dialysis machines with central dialysis fluid delivery (CDDS) system and the conventional dialysis machines, and to analyze the clinical application advantages of the dialysis machines with CDDS system. Methods From August 2017 to March 2018, the bacteria number and endotoxin level at the inlet and outlet sites were compared between the dialysis machines with CDDS system (5 sets) and the Berengdejia dialyzer (5 sets) in the PLA General Hospital Dialysis Center. A retrospective analysis was performed on the incidence of water, circuit and other faults occurred in the dialysis machines with CDDS system (18 sets) and the Berengdejia dialyzer (18 sets) during the period from March 2017 to March 2019. Results The bacteria number and endotoxin level at the inlet were lower in the dialysis machines with CDDS system than in the conventional dialysis machines (t=11.215 and 6.468 respectively; P<0.001). The bacteria number and endotoxin level at the outlet were also lower in the dialysis machines with CDDS system than in the conventional dialysis machines (t=9.154 and 5.490 respectively; P<0.001). The incidence of water, circuit and other faults were lower in the dialysis machines with CDDS system than in the conventional dialysis machines (t=9.147, P=0.010). Conclusion CDDS system can improve the cleanliness of dialysate and reduce the failure rate of dialysis machines during clinical applicati

参考文献

[1]王苏湘, 覃英, 凤灵, 等.透析液配制的检测方法及质量控制研究[J].医疗卫生装备, 2016, 37(11):9-11
[2] 王质刚, 血液净化设备工程与临床[M].人民军医出版社, 20060-94.
[3]涂荣, 李倩.中日血液透析模式比较[J].中国医疗设备, 2014, 29(11):81-83
[4] 川西秀樹.透析液水質基準と血液浄化器性能評価基準2008[J].日本透析医学会雑誌, 2008, 41(3):159-167
[5]焦艳, 白英, 王洪霞, 等.透析用水处理供水系统热消毒与化学消毒的临床比较[J].中国血液净化, 2012, 11(2):102-103
[6] 李松梅.血液透析机的常见故障分析及维修方法[J].中国卫生产业, 2017, 32:19-20
[7] Kawanishi H.全自動透析装置と透析液清浄化[J].Jinko Zoki, 2010, 39:-
[8] Kawanishi H, 土谷, 晋一郎.全自動透析システムの効果と問題点 [J].第回日本透析医学会, 2007, 40:-
文章导航

/

[an error occurred while processing this directive]