The analysis of balloon pressure during percutaneous transluminal angioplasty for hemodialysis access stenosis#br#

  • SUN Hao ,
  • HU Liang-Zhu ,
  • LIU Yang-Dong ,
  • ZHAO Xia ,
  • FU Qi-Ning ,
  • ZHAO Yu
Expand
  • 1Department of General Surgery, Jinshan Hospital, the First Affiliated Hospital of Chongqing Medical University, Chongqing
    401122, China;  2Department of Vascular Surgery, Pinghu Hospital of Shenzhen University, Shenzhen 518111, China;  3Department of Vascular Surgery, the First Affiliated Hospital of Chongqing Medical University, Chongqing 400016, China

Received date: 2021-04-20

  Revised date: 2021-06-07

  Online published: 2021-09-12

摘要

【摘要】目的探讨球囊扩张治疗血液透析血管通路不同位置狭窄时所需的压力差别。方法选择2020 年4 月~9 月于重庆医科大学附属第一医院金山血液透析通路中心就诊并行球囊扩张成形术的血液透析通路患者,收集基本信息、手术资料,并分析不同位置所需的压力区别。结果共259 例患者治疗300 例狭窄病变,男性102 例,女性157 例,自体动静脉内瘘242 例病变,人工血管动静脉内瘘58 例病变。临床成功率100%(300/300),手术成功率98.7%(296/300),手术并发症发生率7.333%,其中血肿形成13例,血管破裂9 例。总平均球囊压力(16.693±6.237)atm,达到充分扩张时球囊压力>14atm 占58.334%(175/300),自体动静脉内瘘与人工血管内瘘之间、人工血管内瘘不同病变位置之间充分扩张所需压力无显著差异,但自体动静脉内瘘近吻合口区域平均球囊压力(18.467± 5.706)atm,高于穿刺区域(t=2.070,P =0.040)和回流静脉(t=4.969,P<0.001)。结论血液透析血管通路狭窄病变对于球囊扩张时达到充分扩张的压力要求更高,不同位置的狭窄所需压力不一,根据这些特点合理选择球囊,方能经济高效地进行腔内介入手术。

本文引用格式

孙浩 , 胡良柱 , 刘杨东 , 赵霞 , 傅麒宁 , 赵渝 . 球囊扩张治疗血液透析血管通路狭窄时所需球囊压力分析[J]. 中国血液净化, 2021 , 20(09) : 624 -627 . DOI: 10.3969/j.issn.1671-4091.2021.09.012

Abstract

【Abstract】Objective To investigate the balloon pressure required during percutaneous transluminal angioplasty (PTA) in the treatment of hemodialysis access stenosis at different locations. Methods Patients with hemodialysis access stenosis and received PTA in Jinshan Hemodialysis Access Center, the First Affiliated Hospital of Chongqing Medical University during April 2020 to September 2020 were enrolled in this study. Demographic and surgical data were collected. The balloon pressure used at different positions was analyzed. Results A total of 259 patients (102 males and 157 females) were enrolled. They had 300 stenosis lesions
including 242 in arteriovenous fistulas and 58 in arteriovenous grafts. The clinical success rate was 100% (300/300), the surgical success rate was 98.7% (296/300), and the incidence of surgical complications was 7.333%, including 13 cases of hematoma and 9 cases of vascular rupture. The total mean balloon pressure was 16.693±6.237 atm, and 58.34% (175/300) of the patients used the pressure of >14 atm for full dilation of the lesions. There was no significant difference in pressure required for full dilation between arteriovenous fistulas and arteriovenous grafts, and among different lesion sites in arteriovenous grafts. However, the
mean balloon pressure for lesions in juxta-anastomotic sites of arteriovenous fistulas was 18.467±5.706 atm, significantly higher than that for lesions in puncture area (t=2.070, P=0.040) and outflow vein (t=4.969, P<0.001). Conclusion The hemodialysis access stenosis requires higher pressure during PTA to achieve full dilation, and the pressure required at different lesion sites is different. According to these characteristics, appropriate selection of balloon makes the endovascular intervention economical and efficient.

参考文献

[1] 金其庄,王玉柱,叶朝阳,等. 中国血液透析用血管通路专家共识(第2版)[J]. 中国血液净化,2019,18(6):365-381.
[2] Tordoir J, Canaud B, Haage P, et al. EBPG on vascular access[J]. Nephrol Dial Transplant, 2007,22(Suppl 2):ii88-ii117.
[3] Aktas A,Bozkurt A,Aktas B,et al.Percutaneous transluminal balloon angioplasty in stenosis of native hemodialysis arteriovenous fistulas: technical success and analysis of factors affecting postprocedural fistula patency[J]. Diagn Interv Radiol,2015,21:160-166.
[4] Swedberg SH, Brown BG, Sigley R, et al. Intimal fibromuscular hyperplasia at the venous anastomosis of PTFE grafts in hemodialysis patients: clinical, immunocytochemical,light and electron microscopic assessment[J]. Circulation,1989,80:1726–1736.
[5] Trerotola SO, Stavropoulos SW, Shlansky-Goldberg R, et al.Hemodialysis-related venous stenosis: treatment with ultrahigh-pressure angioplasty balloons[J].Radiology 2004; 231:259 –262
[6] Vesely TM , Pilgram TK . Angioplasty Balloon Inflation Pressures during Treatment of Hemodialysis Graft–related Stenoses[J]. J Vasc Interv Radiol, 2006, 17(4):623-628.
[7] 赵意平, 郭相江, 施娅雪,等. 高压球囊治疗人工血管动静脉内瘘狭窄的初步体会[J]. 中国血液净化, 2015, 14(001):25-28.
[8] Agarwal SK,Nadkarni GN,Yacoub R,et al. Comparison of Cutting Balloon Angioplasty and Percutaneous Balloon Angioplasty of Arteriovenous Fistula Stenosis:A Meta-Analysis and Systematic Review of Randomized Clinical Trials [J]. J Interv Cardiol,2015,28(3):288-295.
[9] Haskal ZJ, Trerotola S, Dolmatch B, et al. Stent graft versus balloon angioplasty for failing dialysis- access grafts[J]. N Engl J Med,2010,11,362(6):494-503
[10] Mohr BA,Sheen AL,Roy-Chaudhury P,et al.Clinical and Economic Benefits of Stent Grafts in Dysfunctional and Thrombosed Hemodialysis Access Graft Circuits in the REVISE Randomized Trial[J].J Vasc Interv Radiol,2019,30( 2) : 203-211
文章导航

/

[an error occurred while processing this directive]