[an error occurred while processing this directive]
临床研究

A randomized controlled clinical trial on safety and effectiveness of PES14HF polyethersulfone high-flux dialyzer in hemodialysis patients

  • ZHANG Wei-ming ,
  • JIANG Geng-ru ,
  • QIAN Jia-qi ,
  • WANG Bing-shun ,
  • ZHU Chun ,
  • WANG Yong-mei ,
  • HUANG Hai-dong ,
  • YAN Yu-cheng ,
  • NI Zhao-hui.
Expand
  • 1Renal Division, Renji Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200001, China; 2Department of Nephrology, Xinhua Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200092, China; 3Department of Biostatistics, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China

Received date: 2010-05-21

  Revised date: 1900-01-01

  Online published: 2010-08-12

Abstract

【Abstract】Objective To evaluate the safety and efficacy of PES14HF polyethersulfone high-flux dialyzer. Methods Seventy-two patients with stable end stage renal disease and receiving hemodialyis 3 times weekly were enrolled in this randomized parallel control study. High-flux hemodialysis was performed for 4h using PES14HF dialyzer (experiment group) or PES-150DS dialyzer (control group), and the study lasted for one week. The clearances of urea, creatinine, phosphate and β2-microglobulin were calculated. Reactions and adverse events were recorded. Results No significantly differences in serum level changes and clearances of urea, creatinine and phosphate were found between the 2 groups. β2-microglobulin clearance was significantly higher in experiment group (44.50±29.57ml/min) than in control group 25.75±11.27ml/min, P<0.05), so did the decrease magnitude of serum β2-microglobulin (38.87±12.48% vs. 29.53±17.71%, P<0.05). Urea clearance and the higher blood urea improved after the treatment in both groups, and the lowering of blood urea was more significant in experiment group (100%) than in control group (88.57%, P<0.05). Conclusion PES14HF dialyzer is efficient and safe for clinical application.

Cite this article

ZHANG Wei-ming , JIANG Geng-ru , QIAN Jia-qi , WANG Bing-shun , ZHU Chun , WANG Yong-mei , HUANG Hai-dong , YAN Yu-cheng , NI Zhao-hui. . A randomized controlled clinical trial on safety and effectiveness of PES14HF polyethersulfone high-flux dialyzer in hemodialysis patients[J]. Chinese Journal of Blood Purification, 2010 , 9(8) : 429 -432 . DOI: 10.3969/j.issn.1671-4091.2010.08.009

Outlines

/

[an error occurred while processing this directive]